Line 29: Line 29:
   - Make sure you are in Advanced mode by toggling the second button from the left in the main menu bar at the top   - Make sure you are in Advanced mode by toggling the second button from the left in the main menu bar at the top
   - On the left side there is a tab called "Acquisition Group Window" where you can set the integration time, averaging etc.   - On the left side there is a tab called "Acquisition Group Window" where you can set the integration time, averaging etc.
-  - In the spectrum graph windows top menu bar you will find some more functions as well as dark and white measurement options shown as light bulbs+  - In the spectrum graph windows top menu bar you will find some more functions including dark and white measurement options shown as light bulbs
   - To load the calibration files first click on the most left "Create new spectroscopy application" button (Ocean View logo ) on the top main menu   - To load the calibration files first click on the most left "Create new spectroscopy application" button (Ocean View logo ) on the top main menu
   - Click on "Absolute Irradiance" in the dialog box that just opened   - Click on "Absolute Irradiance" in the dialog box that just opened
Line 35: Line 35:
   - Next, in the "Main Controls" section, make sure to disable "Electric Dark" and "Boxcar width" and hit next   - Next, in the "Main Controls" section, make sure to disable "Electric Dark" and "Boxcar width" and hit next
   - Select "Get Irradiance Calibration from File" then hit next   - Select "Get Irradiance Calibration from File" then hit next
-  - Now we have to take a Dark Current measurement. Close the fiber optic cable end with its cover. Additionally, it is recommended to place the closed end somewhere dark during the measurement since the blue cover does let some light through. Hit the light bulb icon to measure and press next when done+  - Now we have to take a Dark Current measurement. Close the fiber optic cable end with its cover. Additionally, it is recommended to place the closed end somewhere dark during the measurement since the blue cover does let some light through.  
 +  - Hit the light bulb icon to measure and press next when done
   - Next, we have to load the calibration file. Hit browse and locate the correct calibration file. Make sure that the calibration file is the right one for the model you are using and that the cable diameter is correct. For more info on the file naming checkout the readme in the github repo: https://github.com/EOLab-HSRW/Ocean_Insight_Spectrometers   - Next, we have to load the calibration file. Hit browse and locate the correct calibration file. Make sure that the calibration file is the right one for the model you are using and that the cable diameter is correct. For more info on the file naming checkout the readme in the github repo: https://github.com/EOLab-HSRW/Ocean_Insight_Spectrometers
   - Press finish and you are done loading the irradiance calibration   - Press finish and you are done loading the irradiance calibration